Field/Laboratory/Industrial Electrochemical Sensors
|
Ion Selective Electrodes :
Ammonium (NH4+), Barium (Ba2+), Bromide (Br-), Cadmium (Cd2+),
Calcium (Ca2+), Carbonate (CO32-), Chloride (Cl-), Cupric (Cu2+),
Cyanide (Cn-), Fluoride (F), Iodide (I-), Lead (Pb2+), Mercury
(Hg2+), Nitrate (NO3-), Perchlorate (CIO4-), Potassium (K+), Silver
(Ag+), Sodium (Na+), Sulphide (S2-), Thiocyanate (SCN-), Water
Hardness,
Do you measure Ammonium, Chloride, Fluoride,
Calcium or Nitrate in water?
Have you used conventional Ion selective
electrodes?
If the answer to any of these questions is yes,
then directION
electrodes are just what you've been looking for!
Main features and benefits include:
- No reference electrode needed
- No filling solution
- Fully submersible and waterproof
- Virtually unbreakable
- Solid state sensors
- Colour coded caps for identification in the
field
- Can be left dry for long periods
- Ideal for unskilled operatives
- Long lifetime
Product Description
directION
electrodes are rugged solid state sensors with built in reference
elements that do not require any filling solution or maintenance.
Fully Submersible
directION
electrodes are waterproof and submersible. This gives the field
operative the ability to cast the probe into the water from awkward
sample points, bridges, banks and boats.
Cable length can be stipulated at the time of
ordering. Maximum length is 10 metres. Standard products are fitted
with 1 metre of cable terminated with a BNC connector.
Laboratory or field use
Though directION
electrodes have a rugged and waterproof construction, there has been
no compromise on technical specification (see table below). Standard
Ion selective performance is surpassed under both laboratory and
field conditions.
What meter can I use?
directION
electrodes can be used with any conventional laboratory or hand held
pH meter with a millivolt display mode. Please stipulate the
connector required on your order. To choose an Ion Selective
Electrode for your specific application,
please see the ISE Selection Chart below :
ORDER
NUMBER |
directION
ELECTRODE |
CONCENTRATION RANGE
(Mol/L) |
LIMITS
(ppm) |
pH RANGE |
TEMP
RANGE
(°C) |
MAIN
INTERFERENCES |
ISAB |
362-75 |
Ammonium (NH4+) |
5 x 10-1 - 5 x 10-5 |
9000 - 0.09 |
1 - 8.6 |
0 - 50 |
K+, Na+ |
CH3COOH |
368-75 |
Barium (Ba2+) |
10-1 - 10-5 |
1.4 - 13,700 |
3 - 10 |
0 - 50 |
Sr2+, K+, Na+ |
CuSO4 |
375-75 |
Bromide (Br-) |
100 - 5 x 10-5 |
0.4 - 80,000 |
1 - 12 |
0 - 50 |
I-, CN-, S2- |
5M KNO3 |
373-75 |
Cadmium (Cd2+) |
10-1 - 1 x 10-5 |
0.1 - 11,200 |
3 - 7 |
0 - 50 |
Hg2+, Ag+, Cu2+
Pb2+, Fe3+ |
5M KNO3 |
361-75 |
Calcium (Ca2+) |
10-1 - 5 x 10-7 |
4,010 - 0.02 |
3.5 - 11 |
0 - 50 |
Ba2+, Al3+, Sr2+ |
KCl |
364-75 |
Chloride (Cl-) |
100 - 3 x 10-6 |
35,500 - 1 |
1 - 12 |
0 - 50 |
I-, Br-, CN-,
S2- |
5M KNO3 |
379-75 |
Cupric (Cu2+) |
100 - 1 x 10-7 |
0.008 - 84,000 |
2 - 7 |
0 - 50 |
Hg2+, Ag+, S2-
Cl-, Br- |
5M KNO3 |
377-75 |
Cyanide (Cn-) |
10-2 - 1 x 10-5 |
0.03 - 260 |
11 - 13 |
0 - 50 |
I-, S2- |
10M NaOH |
365-75 |
Fluoride (F) |
10-1 - 1 x 10-8 |
1,900 - 0.02 |
4 - 8 |
0 - 50 |
OH- |
TISAB |
376-75 |
Iodide (I-) |
100 - 5 x 10-7 |
0.06 - 127,000 |
2 - 12 |
0 - 50 |
CN-, S2- |
5M KNO3 |
372-75 |
Lead (Pb2+) |
10-1 - 1 x 10-5 |
0.2 - 20,800 |
3 - 7 |
0 - 50 |
S2-, Hg2+, Ag+
Cu2+, Fe3+, Cd2+ |
5M NaCIO4 |
374-75 |
Mercury (Hg2+) |
100 - 1 x 10-6 |
0.2 - 201,000 |
0 - 2 |
0 - 50 |
Ag+, S2- |
5M NaCIO4 |
360-75 |
Nitrate (NO3-) |
100 - 7 x 10-6 |
62,000 - 0.4 |
2 - 11 |
0 - 50 |
CI-, NO2- |
4M (NH4)2SO4 |
367-75 |
Perchlorate (CIO4-) |
100 - 2 x 10-1 |
0.2 - 99,500 |
0 - 11 |
0 - 50 |
I-, SCN-, NO3- |
CH3COONa |
366-75 |
Potassium (K+) |
100 - 10-5 |
0.04 - 39,000 |
1 - 9 |
0 - 50 |
Cs+, NH4+ |
4M KCl |
371-75 |
Silver (Ag+) |
100 - 1 x 10-7 |
0.01 - 107,900 |
1 - 9 |
0 - 50 |
S2-, Hg2+ |
5M KNO3 |
315-77 |
Sodium (Na+) |
Sat - 10-6 |
1 ppb |
9 - 12 |
0 - 50 |
Ag+, NH4+,
Li+, K+ |
SISAB |
378-75 |
Sulphide (S2-) |
100 - 1 x 10-7 |
0.003 - 32,000 |
13 - 14 |
0 - 50 |
Ag+, Hg2+ |
10M NaOH |
380-75 |
Thiocyanate (SCN-) |
10-1 - 2 x 10-6 |
1 - 5,800 |
2 - 12 |
0 - 50 |
I-, Cl-, S2- |
5M KNO3 |
370-75 |
Water
Hardness |
2 x 10-1 - 5 x 10-5 |
- |
4.5 - 10 |
0 - 50 |
Ba2+, Sr2+, Cd2+
Cu2+, Na+, K+, NH4+ |
LiAC |
Click Here to go to Electrochemical Sensors Home Page |